Login
Toggle sidebar

Lint: grids

Biographical Date Spacing
error

"Two American radiologists independently of Bucky also came up with the idea of a moving grid. These were Eugene W Caldwell (1870-1918) 6, a radiologist and qualified engineer, who received a patent for an automated timing device to move the grid 4; and Hollis E Potter (1880-1964) 5, who was the first to present the development at a scientific event, namely the winter meeting of the Central Section of the American Roentgen Ray Society (ARRS) in February 1915 4."

Line 55:279 · In a biographical citation, the date should be correctly spaced: '<strong>Hollis E Potter </strong>(1880'
Strong
warning

"The working ability of a grid is described by the grid ratio, which is the ratio of the height of the lead strips (h) to the distance between two strips, i.e. the interspace (D) 7. A grid ratio of 5:1 is generally used for 20-40kVp (i.e. mammography), 8:1 is used for 70-90 kVp technique and 12:1 is used for >90 kVp technique. Higher grid ratios are more effective at reducing scatter, however are more expensive and require a higher dose."

Line 4:54 · Generally, don't use bold in text: '<strong>grid ratio</strong>'

"The strip line density (also known as grid frequency) (number of strips per cm) is 1/(D+d), where d is the thickness of the strip. This is typically 30-80 strips (or grid lines) per cm. Low frequency is 40 to 50 strips per cm, medium frequency is 50 to 60 strips per cm, and greater than 60 strips per cm is high frequency 7."

Line 5:8 · Generally, don't use bold in text: '<strong>strip line density </strong>'

"The strip line density (also known as grid frequency) (number of strips per cm) is 1/(D+d), where d is the thickness of the strip. This is typically 30-80 strips (or grid lines) per cm. Low frequency is 40 to 50 strips per cm, medium frequency is 50 to 60 strips per cm, and greater than 60 strips per cm is high frequency 7."

Line 5:59 · Generally, don't use bold in text: '<strong>grid frequency</strong>'

"Primary transmission is the percentage of primary radiation transmitted through the grid. Ideally, a grid should have 100% primary beam transmission while blocking all the scatter radiation. The primary transmission can be calculated by measuring the ratio of intensity of radiation with a grid over the intensity of radiation without a grid as shown below 8:"

Line 6:4 · Generally, don't use bold in text: '<strong>Primary transmission</strong>'

"The Bucky factor (also known as grid factor) 7,8 is the ratio of incident radiation falling the grid to the transmitted radiation. Although Bucky factor is similar to the primary transmission as described above; there is one major difference: Bucky factor measures both primary and secondary (scattered) radiation transmission, thus making it more practical to determine the additional amount of patient's exposure needed when switching from non-grid to grid technique in order to maintain the same image quality. The formula for Bucky/grid factor is 8:"

Line 18:11 · Generally, don't use bold in text: '<strong>Bucky factor</strong>'

"The Bucky factor (also known as grid factor) 7,8 is the ratio of incident radiation falling the grid to the transmitted radiation. Although Bucky factor is similar to the primary transmission as described above; there is one major difference: Bucky factor measures both primary and secondary (scattered) radiation transmission, thus making it more practical to determine the additional amount of patient's exposure needed when switching from non-grid to grid technique in order to maintain the same image quality. The formula for Bucky/grid factor is 8:"

Line 18:44 · Generally, don't use bold in text: '<strong> </strong>(also known as'

"The Bucky factor (also known as grid factor) 7,8 is the ratio of incident radiation falling the grid to the transmitted radiation. Although Bucky factor is similar to the primary transmission as described above; there is one major difference: Bucky factor measures both primary and secondary (scattered) radiation transmission, thus making it more practical to determine the additional amount of patient's exposure needed when switching from non-grid to grid technique in order to maintain the same image quality. The formula for Bucky/grid factor is 8:"

Line 18:77 · Generally, don't use bold in text: '<strong>grid factor</strong>'

"The contrast improvement factor (K) is the ratio between the contrast with a grid and without a grid. The value of K typically range from 2 to 4 7. The formula is 8:"

Line 22:11 · Generally, don't use bold in text: '<strong>contrast improvement factor</strong>'

"Grid selectivity index measures the ability of a grid to transmit primary radiation while filtering out the scattered radiation, as given by the formula below:"

Line 25:4 · Generally, don't use bold in text: '<strong>Grid selectivity</strong>'
Colons
warning

"The Bucky factor (also known as grid factor) 7,8 is the ratio of incident radiation falling the grid to the transmitted radiation. Although Bucky factor is similar to the primary transmission as described above; there is one major difference: Bucky factor measures both primary and secondary (scattered) radiation transmission, thus making it more practical to determine the additional amount of patient's exposure needed when switching from non-grid to grid technique in order to maintain the same image quality. The formula for Bucky/grid factor is 8:"

Line 18:314 · The first word after a colon should almost always be lowercase. In this case, it's not: ': Bucky factor'.
Acronyms
warning

"RGU"

Line 46:8 · 'RGU' has no definition. Spell it out if it's unfamiliar to the audience.
Headings Valid
warning

Expected headings

  • H1 Terminology
  • H1 Usage
  • H1 Epidemiology
  • H2 Risk factors
  • H2 Associations
  • H1 Clinical presentation
  • H2 Complications
  • H1 Diagnosis
  • H2 Diagnostic criteria
  • H2 Diagnostic clues
  • H1 Pathology
  • H2 Aetiology
  • H2 Location
  • H2 Classification
  • H2 Macroscopic appearance
  • H2 Microscopic appearance
  • H2 Immunophenotype
  • H2 Markers
  • H2 Genetics
  • H1 Radiographic features
  • H2 Plain radiograph
  • H2 Mammography
  • H2 Antenatal ultrasound
  • H2 Transoesophageal echocardiography
  • H2 Ultrasound
  • H2 CT
  • H3 Dual-energy CT
  • H2 Angiography (DSA)
  • H2 MRI
  • H2 CT/MRI
  • H2 Nuclear medicine
  • H3 PET-CT
  • H3 PET-MRI
  • H1 Radiology report
  • H1 Treatment and prognosis
  • H2 Complications
  • H1 History and etymology
  • H1 Differential diagnosis
  • H2 Clinical differential diagnosis
  • H1 Practical points
  • H1 See also

"Characteristics"

Line 3:1 · "Characteristics" is not a recognised heading for this article type.

"Types"

Line 28:1 · "Types" is not a recognised heading for this article type.

"Uses"

Line 35:1 · "Uses" is not a recognised heading for this article type.
Parentheses
suggestion

"The strip line density (also known as grid frequency) (number of strips per cm) is 1/(D+d), where d is the thickness of the strip. This is typically 30-80 strips (or grid lines) per cm. Low frequency is 40 to 50 strips per cm, medium frequency is 50 to 60 strips per cm, and greater than 60 strips per cm is high frequency 7."

Line 5:44 · Use parentheses judiciously. There are at least 3 sets in this paragraph.
Semicolons
suggestion

"The Bucky factor (also known as grid factor) 7,8 is the ratio of incident radiation falling the grid to the transmitted radiation. Although Bucky factor is similar to the primary transmission as described above; there is one major difference: Bucky factor measures both primary and secondary (scattered) radiation transmission, thus making it more practical to determine the additional amount of patient's exposure needed when switching from non-grid to grid technique in order to maintain the same image quality. The formula for Bucky/grid factor is 8:"

Line 18:283 · Use semicolons judiciously.

"virtual grid: no actual grid is used; latest innovation for scatter reduction by digitally reconstructing a radiograph"

Line 30:51 · Use semicolons judiciously.

"moving grids (also known as Potter-Bucky or reciprocating grids): eliminates the fine grid lines that may appear on the image when focused or parallel grids are used; cannot be used for portable films"

Line 33:173 · Use semicolons judiciously.

"Two American radiologists independently of Bucky also came up with the idea of a moving grid. These were Eugene W Caldwell (1870-1918) 6, a radiologist and qualified engineer, who received a patent for an automated timing device to move the grid 4; and Hollis E Potter (1880-1964) 5, who was the first to present the development at a scientific event, namely the winter meeting of the Central Section of the American Roentgen Ray Society (ARRS) in February 1915 4."

Line 55:273 · Use semicolons judiciously.