"The term detector pitch is the table distance travelled in one 360° gantry rotation divided by the beam collimation 2."
"Beam pitch is defined as table distance travelled in one 360° gantry rotation divided by the total thickness of all simultaneously acquired slices ref."
"P = 1.0: x-ray beams are contiguous for adjacent rotations"
"P >1.0: x-ray beams are not contiguous for adjacent rotations; i.e. there are gaps in the x-ray helix, but the full volume is still irradiated, only with fewer projections per rotation"