"This technique is sometimes referred to, perhaps more accurately, as dynamic susceptibility contrast-enhanced MR perfusion, still abbreviated to DSC. This should not be confused with dynamic contrast-enhanced (DCE) MR perfusion, which relies on T1 shortening due to gadolinium-based contrast."
"can be reduced with shorter TE (not shorter than 30ms at 3T)"
"exacerbated by long TR"
"can be reduced with shorter TR ("
"B1 intensity variations"
Expected headings
"Physics and technique"
"Pitfalls"
"Artifacts"
"Dynamic susceptibility contrast (DSC) MR perfusion is one of the most frequently used techniques for MRI perfusion, and relies on the susceptibility induced signal loss on T2*-weighted sequences which results from a bolus of gadolinium-based contrast passing through a capillary bed. The most commonly calculated parameters are rCBV, rCBF, MTT and Tmax 5."